Difference between revisions of "Lip(Sys)2/Near-infrared Spectroscopy"

From DAAP
Jump to navigation Jump to search
m (Text replacement - "daap.eu" to "daap.dsi.universite-paris-saclay.fr")
 
(12 intermediate revisions by 3 users not shown)
Line 1: Line 1:
{{Under construction}}
 
 
 
{{#invoke:Device|infobox
 
{{#invoke:Device|infobox
 
| Title = Büchi NirFlex N500  
 
| Title = Büchi NirFlex N500  
Line 7: Line 5:
 
| Website =  
 
| Website =  
 
| Keywords = FT-NIR, NearInfrared, SPIR, ProcheInfrarouge
 
| Keywords = FT-NIR, NearInfrared, SPIR, ProcheInfrarouge
| Picture =  
+
| Picture = http://daap.dsi.universite-paris-saclay.fr/w/images/d/d0/Nirflex_N500_low_res.jpg
| Latitude =  
+
| Latitude = 48.7618867
| Longitude =
+
| Longitude = 2.273395
 
}}
 
}}
 
+
'''Fourier transform infrared spectroscopy, NIRFlex N-500, Büchi'''
== Büchi NirFlex N500 ==
+
== Description==  
  
 
The Büchi Nirflex N500 is a Fourier Transform Near-Infrared spectrometer suited for the analysis of solids by diffuse reflectance.  
 
The Büchi Nirflex N500 is a Fourier Transform Near-Infrared spectrometer suited for the analysis of solids by diffuse reflectance.  
Line 23: Line 21:
 
- XL vials (e.g. a 20 ml glass scintillation vial or larger),  plastic bags (single sample)
 
- XL vials (e.g. a 20 ml glass scintillation vial or larger),  plastic bags (single sample)
 
The illumination spot diameter is 9 mm for Petri plates and XL Vials sample holders and 8 mm for 2mL vials and Tablets holders.
 
The illumination spot diameter is 9 mm for Petri plates and XL Vials sample holders and 8 mm for 2mL vials and Tablets holders.
 +
 +
 +
[[File:Nirflex N500 sample holders.jpg]]

Latest revision as of 14:26, 26 April 2021

Device
Nirflex_N500_low_res.jpg

Pilot process
Keywords
Organisation
Latitude
48.7618867
Longitude
2.273395

Fourier transform infrared spectroscopy, NIRFlex N-500, Büchi

Description

The Büchi Nirflex N500 is a Fourier Transform Near-Infrared spectrometer suited for the analysis of solids by diffuse reflectance. The spectral range is from 800 to 2500 nm (12 500 ~4000 cm-1) with an 8cm-1 resolution. The scan speed is 2 scans/second. Spectral data can be processed by the proprietary Nircal® chemometric software (PCA, SIMCA, PLS algorithms) or exported to the JCAMP-DX format for further data processing. Four adapters are available to accommodate: - 2mL vials such as HPLC autosamplers vials (6 positions) - Petri plates for bulk solids (single sample) - Tablets for direct analysis of tablets (round or oval shaped) and capsules (10 pos.) - XL vials (e.g. a 20 ml glass scintillation vial or larger), plastic bags (single sample) The illumination spot diameter is 9 mm for Petri plates and XL Vials sample holders and 8 mm for 2mL vials and Tablets holders.


Nirflex N500 sample holders.jpg